Sirrus Clinical Chemistry Analyzer

Sirrus automatically reads and manages bar coded reagent bottles and sample tubes. Most reagents have 30-day on-board stability to minimize waste. The intuitive Windows-based software screens make learning and operating Sirrus easy and quick.
 
Sirrus can run at a blazing 400 tests per hour. Dual reagent probes and a dedicated sample probe translate into higher throughput and faster patient turn-around time. The increased efficiency and use of reusable cuvettes translate into lower costs and thus higher profits per test.
 
Features and Benefits
 
Reagent System
  • 36 positions for either R1 or R1/R2 reagents
  • On-board reagent cooling temperature: 10 ± 2° C
  • Liquid, ready-to-use reagents
  • 30-day on-board stability for most reagents
  • 30-day calibration stability for most reagents
  • Reagent probe with automatic level detection
  • Automatic inside and outside probe washing after each reagent addition
Sampling Unit
  • Sample Type: Serum, plasma, urine, CSF, hemolysate
  • 45 patient positions on sample tray (30 primary tubes) including 5 STAT positions
  • Standard cup and blood collection tubes accepted
  • Integral bar code reader for primary tubes
  • Sample volume: 3-30 μL
  • Auto-dilution capability
  • Auto-rerun capability
  • STAT interruption permitted during analysis
  • Sample probe with automatic level detection
  • Automatic inside and outside probe washing after each sample addition
Reaction Unit
  • 60 semi-permanent plastic cuvettes
  • 37° ± 0.1° C reaction temperature
  • Automatic 10-step cuvette washing system
  • Cuvette optical length: 8 mm
Optical Absorption Measuring Unit
  • Direct measurement of reaction cuvette
  • Monochromatic and bichromatic measurement
  • Wavelength: (12 discrete) 340, 380, 405, 450, 
  • 505, 546, 570, 600, 660, 700, 750, 800 nm
  • Cell blank reading corrected after each cuvette washing
  • Tungsten halogen lamp, 2000 hours
  • Diffraction spectrophotometer with concave grating
  • Silicon photo-diode detectors
Ion Selective Electrode (ISE) Unit
  • Na, K, Cl test items
  • Total of 45 seconds test time to perform all three tests
  • Low sample volume: 60 μL
  • 24-hour ISE calibration stability
Technical Specifications
  • Dimensions: 31.5" W x 25.2" D x 20.5" H
  • Weight: approx. 210 pounds
  • Power Supply: AC 115V 50/60Hz
  • Rated Power Consumption: 600VA
  • Water supply/drain: 3.5 liters/hour de-ionized water

Source:SIRRUS